SPEEDCAL

HOME arrow SPEEDCAL arrow REFERENCES
DATA SHEET SpeedCal PDF Print
 DIMENSIONS AND WEIGHT OF SpeedCal WITH FRAME RACK

 height: 1780 mm  width: 846 mm  depth: 900 mm  weight: 120 kg

 DIMENSIONS AND WEIGHT OF SpeedCal FOR INSTALLATION ON A TABLE

 height: 1170 mm  width: 846 mm  depth: 900 mm  weight: 110 kg

 ELECTRICAL DATA

operating voltage                   100-240 V / 50 Hz oder 60 Hz
protection class IP54

 APPLICATION

  • multi-channel piston pipettes
  • nominal volume: 10 µl to 1,2 ml
  • optional supplementary equipment for single-channel pipettes

 WEIGHING CELLS

typeSartorius XX 40-0002 
number12
resolution 0,01 mg 
range of taring 12 g 

repeatability

< 0,02 mg
(as standard deviation)   
error of linearity < 0,02 mg
error of measurement (k = 1) 

< 0,02 mg (in a range of 0 to 1 g ~ 0 to 1000 µl)

settling time (typical) ca. 10 s 
sensitivity to drift of temperature between + 10 °C and + 30 °C <  ± 1 x 10 -6 ppm/K

 WEIGHING TUBES (removable, able to be dismantled)

distance and volumedistance: 9 mm; volume: 12 ml

 CONFORMITY

  •  according to DIN EN ISO 8655

testing time10 minutes for the complete test of 3 adjusted
 volumes, is equivalent to 360 singular values
 (operating time of the unit),
 plus manual pipetting 

 REQUIREMENTS TO THE INSTALLATION LOCATION

  • level and flat
  • free of mechanical vibrations
  • clean and dust-free
  • sources of strong electromagnetic interferences should be avoided
  • infiltration (e.g. by air conditioning systems) and direct solar radiation should be avoided

allowed ambient temperatureconstant ± 1 °C between + 10 °C to + 30 °C 

 DATA LOGGING

  • standard-computer (optional: Notebook), monitor, printer
  • operating system: Windows 98, Windows 2000, Windows NT

 CE-CERTIFICATION

  • according to guidelines 73/23/EWG
  • according to guidelines 89/336/EWG
  • according to guidelines 90/384/EWG (optional)

 WARRANTY

  •  12 months
 
< Prev   Next >